Lab 9.
Procedure.
You will collect data with a lab partner (who
should be noted in your write-up). However, all the data analysis
and write-up should be done without collaboration.
Before the experiment
You should have
- gas cell with KBr windows
(pathlength=10 cm)
- concentrated H2SO4
- D2O
- CO2 tank
- C2H2 lecture
bottle
- C2D2 lecture
bottle
- vacuum line filling apparatus
- FTIR spectrometer Spectrum One
from Perkin-Elmer
Inc. The local version of the manual for Spectrum One is available here
(courtesy of K.Burke).
During the experiment
Part A. CO2
In this experiment, several
vibrational-rotational infrared bands of CO2 will be
recorded at medium-to-high resolution (ca. l cm-1).
These spectra will be analyzed to extract information about hot
bands.
- Pump out the system for 10-15
minutes (have IR cell attached with its valve open).
- Close off the line to the cell.
- Take your background spectrum. Use
the highest resolution (0.4 cm-1 if
possible), set the signal gain to 1
and the number of scans to 16 or greater.
- Return cell to apparatus. Add
approximately 1atm of CO2.
Record IR absorption spectrum. Two spectral ranges
are of interest: 550-750 cm-1 and 2250-2450 cm-1.
At this high pressure you should observe that absorption
extends quite far from the maxima due to rotational
unresolved P and R branches.
- Repeat the same with 150 Torr of
CO2. In the region of bending mode absorption,
besides seeing unresolved rotational P and R
branches, you should see sharp lines at 618, 648, 720 cm-1
and maybe more. These lines are due to hot bands.
- Repeat the same with a piece of dry ice. Did the hot
bands disappear?
Part B. HCl and DCl
In this experiment rotational fine structure of
the infrared vibrational spectrum of HCl and DCl will be recorded
at medium-to-high resolution (ca. l cm-1). These
spectra will be analyzed to extract rotational constants and
obtain the moment of inertia of the molecule and thus the
internuclear separation. The pure vibrational frequencies will
also be measured.
- HCl and DCl can be synthesized by addition
of D2O to sulfuric acid (1:1) and poring it
dropwise over solid NaCl or KCl. As a result, gaseous
mixture of DCl/HC will be obtained: D2SO4(l)
+ 2NaCl(s) -> 2DCl(g) + Na2SO4(s).
The trick is to collect water vapor before the IR cell so
that it won't damage the windows. Use a trap cooled by
dry ice for that.
- Pump out the system for 5-10
minutes (have IR cell attached with its valve open).
There should also be a dewar of liquid nitrogen on the
gas trap leading to the pump; this will condense out the
HCl and DCl so these do not go through the pump and into
the air.
- Close off the line to the cell.
- Take your background spectrum. Use
the highest resolution (0.3 cm-1 if
possible), set the signal gain to 1
and the number of scans to 16 or greater.
- Return cell to the vacuum system.
Open its valve connected to the line with HCl/DCl. You
should fill to about 200 Torr (mm Hg) of the mixture.
- You may want the TA to assist here
in filling the cell until you are comfortable doing it.
- Close off the IR cell valves,
remove itl and take spectrum using the same parameters as
for background spectrum.
- Repeat procedure for other gases.
- When completely finished, turn off
the pump, remove the gas trap and put it in a fume hood,
and return the IR cell to a desiccator.
For HCl, you are interested in the
region around 2900 cm-1. For DCl, the region of
interest is around 2100 cm-1. Use the peak report
function to create a table of the wavenumber position and
absorbance of each peak within these regions.
Part C (if time allows).
In this experiment, several
vibrational-rotational infrared bands of C2H2
and C2D2 will be recorded at medium-to-high
resolution (ca. l cm-1). These spectra will be
analyzed to extract rotational constants for use in the
calculation of accurate values for the C-H and C-C bond lengths.
The role of symmetry and nuclear spin in determining the
activities and intensity patterns of the spectral transitions is
also examined. From such considerations, the infrared bands can
be assigned to specific modes of vibration and values can be
deduced for the fundamental vibrational frequencies of C2H2
and C2D2 [1,2]
TABLE 2
Infreared regions to be scanned for acetylenes
| |
Frequency (cm-1) |
| |
C2H2 |
C2D2 |
| Scans at ~ 300 Torr |
| Survey scan |
400 - 4000 |
400 - 4000 |
| n2
- n5 |
1235 - 1255 |
1220 - 1235 |
| n1
- n5 |
2635 - 2650 |
2160 - 2175 |
| n3
- n4 |
2675 - 2690 |
1920 - 1935 |
| Scans at ~ 25 Torr |
| Survey scan |
400 - 4000 |
400 - 4000 |
| n5 |
720 - 740 |
530 - 550 |
| n4
+ n5 |
1235 - 1255 |
980 -1120 |
| n3 |
3275 - 3325 |
2425 - 2450 |
- Pump out the system for 10-15
minutes (have IR cell attached with its valve open).
- Close off the line to the cell.
- Take your background spectrum. Use
the highest resolution (0.3 cm-1 if
possible), set the signal gain to 1
and the number of scans to 16 or greater.
- Return cell to apparatus. Add
approximately 300 Torr of C2H2
for a survey scan. Table 2 indicates the spectral regions of
interest and the approximate pressures that give
satisfactory intensities. These pressures may require
some adjustment depending upon the resolution
capabilities of the instrument since the peak absorbance
of a narrow line increases as the spectral resolution
improves. For the survey scan, a resolution of 4 cm-1
is adequate to permit rapid recording at reasonable
signal-to-noise ratio. The regions to be studied in
detail should be scanned more slowly at an expanded scale
to permit accurate frequency measurements. A resolution
of at least 1.5 cm-1 is needed to resolve the
rotational structure of the acetylene bands, and a value
of 0.5 cm-1 or better is desirable. Detailed
instructions for operating the spectrometer will be given
in the laboratory.
- Reduced the cell pressure to about 25 Torr
so that the strong n3 and n5
fundamentals and the n4 + n5
combination band have a more reasonable intensity.
Expanded scans of the latter bands are recorded according
to Table 2, preceded by a second survey scan.
- We plan to have it already done
for you, but just in case - C2D2
can be synthesized by addition of D2O to
calcium carbide using the apparatus shown in SGN. About
2.5 g of calcium carbide is placed in flask F which is
then evacuated to remove traces of H2O; 0.5 mL
D2O is added with a syringe to flask F through
the rubber septum and the entire system up to the vacuum
stopcock V is allowed to "cure" at room
temperature for about 5 to 10 min to allow deuterium
exchange with H2Oadsorbed on the walls of the
system. The pressure should be monitored and kept below 1
atm during this period, reducing it if necessary by
opening and closing V. The system is then evacuated,
after which the cold traps are put in place and 0.5 mL D2O
is added to flask F. The pressure will rise and then drop
as the C2D2 is condensed in the
storage vessel cooled by liquid nitrogen. When the
pressure drops to a few Torr, another increment of D2O
is added and the procedure is repeated until a
total of 4 mL has been added. When gas evolution has
slowed or stopped, stopcock A is closed and the reaction
flask and water trap are placed in the hood, open to
allow any further reaction to occur harmlessly.
- Fill the infrared cell with C2D2
by evacuated the system with the liquid nitrogen trap
still in place. Stopcock V is then closed and the
nitrogen Dewar is lowered to allow the storage vessel to
warm slowly until the pressure is about 300 Torr. The
stopcocks C and C' leading to the infrared cell are then
closed and, as warming continues, the system pressure is
monitored and adjusted with stopcock V if necessary to
keep it below 1 atm. When room temperature is reached,
stopcock B' can be closed to save some residual C2D2
as "insurance" until all cell pressure
adjustments and spectral measurements are completed. Do
not open stopcock B' to air with the storage vessel in
liquid nitrogen since liquid oxygen will condense on top
of the acetylene, forming a potentially explosive
mixture. At the end of the experiment, the
acetylene in the infrared cell and in the storage vessel
can be disposed of by simply exhausting it through the
roughing pump of the vacuum system.
- Repeat measurements the same way as with C2H2
{Commercial acetylene is widely used for
welding purposes and is shipped dissolved in acetone, in which it
is extremely soluble. The acetone is retained by a porous filler
material within the cylinder so that the discharged acetylene is
typically >99%. If desired, residual traces of acetone can be
eliminated by passage through a Dry Ice/isopropanol trap. In its
free state, acetylene may decompose violently; the stability
decreases at higher pressures. At pressures below 1 atm, the
sampling conditions of this experiment, the gas can be handled
safely but one should, of course, wear safety glasses and
exercise reasonable judgement. Unalloyed copper, silver, and
mercury should never be used in direct contact with acetylene,
particularly when wet, owing to the possible formation of
explosive acetylides.}
{A higher D/H ratio can be achieved by
adding the acetylene to a storage bulb containing 5 mL of 99 + %
D2O and about 1 g of basic alumina.
The latter serves to promote the exchange of acidic protons on C2H2
with the D2O and thereby improves
the D content of the acetylene. For best results, the protons on
the basic alumina should first be exchanged by adding a few mL of
D2O and evacuating the storage bulb
prior to addition of more D2O and
acetylene. After exchange at room temperature for a few hours,
the storage bulb is cooled in a Dry Ice/isopropanol bath and the
enriched C2D2
is distilled into the infrared cell.}
After you finish
- turn off the pump and disconnect it from the vacuum line
(let the air into the pump)
- clean up the mess after yourself
- show you notebook and the apparatus to the TA or the
instructor before you leave
Tentatively, plan to run CO2 during first day,
HCl/DCl on a second day and acetylenes on a third day.
Last updated on 02/12/07