Lab 6.
Procedure.
Parts of this lab are adapted from Schwenz and Moore1 and analogous experiment at Dartmouth College2.
You will collect data with a lab partner (who
should be noted in your write-up). However, all the data analysis
and write-up should be done without collaboration.
In this lab you will study quenching of Ru(bpy)32+phosphorescence
by time resolved and steady state measurements of its
luminescence. You will compare the two techniques and identify
different mechanisms of quenching. Following quenchers will be
used: O2, Fe(H2O)63+, Fe(H2O)62+ , Fe(CN)63- , and
Fe(CN)64- . You and your partner will be given
much freedom in the design of the experiment. Please make sure to
plot a strategy among yourselves before coming to the laboratory.
This simple act of communication and preparation will save
valuable time in the laboratory.
Before the experiment
Record room temperature
- Section A. You should have a few glass
pipetts, a 60 mL plastic syringe with a narrow teflon
tubing approximately 6 cm long and the two gas tanks -
with oxygen and nitrogen, both equipped with a regulator
and a purging device as in the picture on the right. The
purpose of this purging device is to have either O2
or N2 at atmospheric pressure by means of a
slow flow of gas through it. Pumping known volume of each
gas will provide a measured mixture of the gases.
- Section B. You should have following
stock solutions available:
- 1.8 x 10-4
M of Ru(bpy)32+
in water
- 0.5 M HCl solution
- 14.8 mM Fe(H2O)63+
(made from FeCl36H2O)
in 0.5 M HCl (in order to prevent the formation
of ferric hydroxide and other dimeric ferric
species)
- 21.8 mM Fe(H2O)62+
(made from FeCl26H2O)
in 0.5 M HCl
- 11.5 mM Fe(CN)63-
(made from K3Fe(CN)6) in water
- 8.8 mM Fe(CN)64-
(made from K4Fe(CN)6) in water
- record a fluorescence profile of *Ru(bpy)32+
and absorption spectra for Fe(H2O)63+
, Fe(H2O)62+ , Fe(CN)63- and
Fe(CN)64-
at low concentrations (so that
absorbance is less than 1 in a 1 cm cuvette, you can use
a plastic cuvette for this)
- By mixing the above stock solutions in appropriate
proportions, prepare following stock solutions (in 25 mL
volumetric flasks) for your measurements (and
label them):
- 9 x 10-5
M of Ru(bpy)32+
in water
- 9 x 10-5
M of Ru(bpy)32+
in 0.25 M HCl
- 9 x 10-5
M of Ru(bpy)32+
+ 1 mM Fe(H2O)63+
in 0.25 M HCl
- 9 x 10-5
M of Ru(bpy)32+
+ 10 mM Fe(H2O)62+
in 0.25 M HCl
- 9 x 10-5
M of Ru(bpy)32+
+ 1 mM Fe(CN)63- in water
- 9 x 10-5
M of Ru(bpy)32+
+ 1 mM Fe(CN)64- in water
- For your Stern-Volmer plot, you will need
to vary the concentration of a quencher. Using your
prelab calculations, make 4 or 5 solutions (~ 5 mL each)
with each quencher at different concentrations suitable
for the Stern-Volmer plot. Since you measuring the
lifetimes, concentration of of Ru(bpy)32+complex
does not have to be constant.
-
During the experiment
The experimental arrangement of equipment is
shown in the Figure below.

The N2-laser (VSL-337 from LSI) delivers 4 ns pulses at
337 nm of about 100 mJ each. This wavelength is shorter than the maximum of
Ru(bpy)32+ absorption but we can always
adjust its concentration to gain absorbance to the desired value
around 1. The triggering photodiode (ET 2000 from EG&G)
initiates recording the measurements by the detecting photodiode
(DET2 from Thor Labs). The signal is collected on a digital oscilloscope (TDS-210 form Tektronix) which is interfaced to a computer. The data from the
oscilloscope will be transferred to a computer via serial port
using Wavestar program, also from Tektronix. You can review the help file for this program before
starting the lab either on the lab computer or by downloading
this file wvstar.hlp (warning, its size is 510 k). You will need to use
probably 250 ns/division timescale and either 5 mV or 10 mV
sensitivity scale on the oscilloscope (correct setup should be
stored as a setup 5: Push Save/Recall|Setup scroll to 5|Recall). For running an experiment you need to:
- turn on the Thor photodiode
- turn on the scope (choose setup 5 as
described above)
- put your sample into the cuvette holder
(you should use a quartz cuvette)
- turn on the laser
- push the RUN button
- when acquisition is complete (takes less than 30 sec),
trasfer the kinetic to the computer
Section A. Quenching by O2
- fill a quartz cuvette approximately 1/2
full with your stock solution of 9 x 10-5 M of *Ru(bpy)32+ in water
and record its luminescence kinetics using
above procedure; store your data; the *Ru(bpy)32+
lifetime in this solution should be shortened
(quenched) in accordance to the natural abundance of O2
in atmosphere
- purge the solution with N2 gas
by slow bubbling via a glass pipett for 5 min and record
the luminescence kinetics; make sure that the solution
has reached the room temperature; the Ru(bpy)32+
lifetime in this solution should correspond to the
unquenched lifetime
- now is a tricky part: you have to make
three or four measurements at intermediate oxygen
concentrations; my suggestion is to use 60 cc syringe
- fill it with a mixture of the two gases by
sucking each gas through the outlet of the
corresponding purging device
- purge the mixture slowly through the solution and
put the cap on the cuvette as soon as you finish
- with the cap on, shake the solution gently for 1
min
- repeat the procedure with the same gas
ratio again at least five times
- after waiting a few minutes for the solution to
warm up, make the lifetime measurements as before
- the highest oxygen concentration will be
achieved by purging the solution with pure O2
gas directly from the tank by slow bubbling via a glass
pipett for 5 min; again, make sure that the solution has
reached the room temperature; record the luminescence
kinetics and measure the lifetime
Section B. Quenching by ions
- record luminescence kinetics of *Ru(bpy)32+
in water and 0.25 M HCl solutions as before, without
deoxygenation; measure the lifetimes - those will be the
lifetimes for the Stern-Volmer evaluation of quenching in
corresponding ionic solutions
- you already, presumably, made 4 or 5
solutions (~ 5 mL each) with each quencher at different
concentrations suitable for the Stern-Volmer plot; record
luminescence kinetics for each of them, measure the
lifetimes
After you finish
- clean up after yourself
- show you notebook and the setup to the TA or the
instructor before you leave
Tentatively, plan to run O2 quenching on a first
day and quenching by ions on the last two days. On a third day,
if you have time, you might progress to your own project. It is
encouraged that you discuss the project with the instructor, so
that appropriate adjustments can be made.
References
- Physical Chemistry: Developing a Dynamic Curriculum,
Ed. R. W. Schwenz and R. J. Moore, 1993, American
Chemical Society, Washington, DC.
- Chem 10 Lab manual, Department of Chemistry, Dartmouth
College
Last updated 01/09/06.