 |
NMR Analysis for 13C
with the Unity 400
|
|
The 13Carbon NMR spectrum of an organic compound
provides information concerning:
- The number of different types
of carbon atoms present in the molecule. Hybridization has a
strong affect: sp3 carbons absorb at 0-100 ppm,
sp2 carbons at 100-220 ppm, sp carbons at 70-100 ppm.
- The electronic environment
of the different types of carbons
- The number of "neighbors"
for a carbon (splitting -- carbon-carbon splittings are not observed:
because 13C abundance is 1.1%, the chance of 2 neighboring
is about 0.01%.)
The major differences between 13C-NMR
in comparison to 1H-NMR spectra include:
- No integration of carbon
spectra
- Wide range (0-200 ppm) of
resonances for common carbon atoms, typical range for protons
1-10 ppm (Carbonyl carbons are particularly characteristic, absorbing
at about 170-220 ppm.)
Before logging on -- Open the login
book and enter the user name along with the date and time. Check to
make certain no one is currently using the instrument, enter the
password for the log-on computer. If the instrument is currently
being used, check with the person to see if it is possible to log them off
and log in. Normally if a user forgets to logoff the
instrument is locked.
Logging in -- Move the mouse to see
the login window. Type the login name and when prompted, the
password.
In the login process, a series of programs start. Wait until the Front Panel us displayed:

The most important applications are on the front panel: vnmr (an icon with a blue spectrum on a
black background), and the File Manager (an icon with an open filing cabinet). To open an application
click on it once. Additional clicks will open multiple copies of VNMR which will not operate.
Click on the vnmr icon on the Front
Panel once. Wait until the Acqi button shows up at the right of the first row of buttons, and the Acquisition
Status-window opens.
Menu-driven operation. The most important
commands can be called by clicking on the buttons in various menus.
However, it does not give access to the full range of commands.
The Main Menu button is on the first row of buttons at any time.
Click on it to retrieve the second row of buttons on the Main
Menu. These buttons are arranged in the normal sequence for obtaining
a plot of a spectrum. In any menu, the buttons on the second row
are defined by clicking the Help button.
Command mode. Type commands in the Input Window. The input window must be selected before typing a command.
Selecting the Workspace -- NMR spectra
are obtained in experiments. An experiment is actually a directory
containing files of the text associate with the sample, the file
with the parameters for acquisition and data processing, the FID
file, and the spectrum file after transformation of the FID. Each
user has his own experiments, which means data in an experiment
will be the same, no matter what the other users have done. The
data in an experiment is lost if the set up command is given,
if another FID is retrieve, by starting an acquisition, or by
delete the experiment. At the first log in there os one experiment. To create more, click Workspace - Create New. Do not create more experiments
that needed at a time. They occupy disk space. Use the Delete
button to erase the unnecessary ones. To see the text associated
with each experiment click Library. To move to (join) experiment
3, click the Exp 3 button, or type jexp3 in the input window.
Setting the Parameters -- From the Main
Menu click Setup. The setup copies standard
parameters file into the parameters file for the current experiment.
First, specify the nucleus and the solvent. For the five solvents
listed as buttons, appropriate lock parameters can be retrieved.
There are buttons defined for proton or carbon in d-chloroform.
All the frequencies are set based on the lock frequency. Select the solvent correctly or the spectral window will be improperly set and some
region of the spectrum will show up folded. Or select the parameters
by typing in the input window, e.g. setup(C13',CDCl3').
To display the parameters at any time, type in the input window dg (display group parameters). Not all the parameters
are listed by dg. To inquire about the
value of a parameter, type the parameter name followed by question
mark, e.g. gain?. To modify the value of
a parameter, type parameter_name=new_value.
Note that text values should be in quotes, e.g. gain=n.
Procedure for 13C -- Follow steps 1 through 12, details
are given below.
1. Remove standard and
Insert sample into the magnet
2. Set up Parameters
3. Retrieve
shims
4. Tune, only
if necessary
5. Manual
lock
6. Manual
shim
7. Now change
parameter nt, then type ga to collect data
8. Process and print out data
9. To Save Click Data, then type svf and file name
10. Plot
Data
11. Remove
Sample and put Lock Standard Back
12. Log off the Computer
1. Remove the sample from the top of the magnet
by holding onto the top of the spinner. Remove the previous sample
from the spinner with a firm pull on the top end and store it
in a safe place. Insert the sample by typing commands e and i.
To Insert a Sample into the Magnet --
Push the sample tube through the spinner using a twisting motion
to avoid breaking the sample tube. The bottom of the sample should
be inserted into the wide end of the
spinner first. Adjust the spinner's position using the gauge.
Wipe the outside of the tube and the spinner with a soft tissue.
When moving the sample/spinner assembly, always hold the top of
the spinner to avoid repositioning of the spinner. Never insert
a sample unless the bore of the probe is known and there is a sample
in the spinner.
Place the sample tube and spinner into the transfer
tube opening in the center of the magnet. Make sure that there
is enough air to support the sample before letting go of the sample
tube and spinner. Type 'i' so that the sample will be lowered
into the probe. Listed for two click sounds (or
a click and then a cluck) which indicate if the sample
tube sits in the right position.
If there is no "click" of the sample
resting in the probe or the spin indicator light does no stay
on (it will blink until the sample is spinning at the desired
rate), try ejecting and inserting again. Cover the transfer tube
opening in the center of the magnet briefly with one hand and
wait for the spinner light to turn on.
Note -- Always leave a sample in the
magnet with the lock on at a level of 30%
to 70%. The purpose of using a lock solvent is to minimize magnetic
field drift.
2. Set up parameters by clicking setup,
then select the nucleus and solvent, or by typing rtp,
then answer carbon 13: C13. If the solvent
is correct then type su.
3. Retrieve shim set by typing rts,
plus the shim file name, such as CDCl3, then su.
Type rts, at the prompt, enter CDCl3, and then su.
This will load a standard shim settings of the using probe. This
is a good start to shim a sample.
Once a good shim setting has been established for a
sample type, it is generally possible to switch between different
nucleus without further shimming.
4. Tune. Only if necessary, details
are given at the end of these instructions.
5. Manual lock . Or try
autolock (type lock) if manual lock is
difficult.
Enter 'acqi' to open the acquisition
window. Turn the lock control off. The
lock signal appears as a sine wave if the lock signal is off-resonance,
and as a dc signal if the lock signal is on-resonance. Adjust
Z0 until the lock signal is on resonance.
Increase the lock power to 20 and the lock gain to 30
or higher until there is sine wave signal. Adjust Z0 to get a dc
signal.
Adjust the lock power and lock gain until the dc signal
is from 30% to 70% of the full screen.
Turn the lock on. Adjust the phase so that the dc signal
is highest at the above setting of the lock power and lock gain.
6. Manual shim. Or try
autoshim (type shim) if manual shim is
difficult.
The screen will show a digital lock amplitude and two
horizontal lock signal bar graphs. The top one shows the total
lock signal, the bottom one shows either 0-20%, 20-40%, 40-60%,
60-80%, or 80-100% of the full scale, using different colors for
each range.
(A) Adjust each of the following
to maximize lock signal- Z1C, Z2C, Z1, Z2. Normally only adjust Z1C and Z2C.
(B) Increase Z1 by 20 units,
maximize lock signal with Z2.
(C) If level is higher
or the same, repeat step (B). If it is lower go to step (D).
(D) Decrease Z1 by 20 units,
maximize lock signal with Z2.
(E) If level is higher
or the same, repeat step (D). If it is lower, increase Z1 by 20
units and
maximize lock signal with Z2 and the sample is shimmed.
Continue the steps until signal is maximized.
7. Now change parameter nt, then type ga to
collect data.
8. Process and print out data.
Type command 'wft aph' if go was used instead of ga. The following are a few possible
commands to further process the spectrum if needed: to display
the scale click Dscale or type dscale if
the scale is in hertz, type axis=p to view the full spectrum in the full screen f full, to
adjust the vertical scale use the middle mouse button directly
on the spectrum or type vs=value wanted, to zoom in on the region
within the cursors left-click on the left and right-click on the
right of a region in the spectrum then click Expand to change
the limits of the spectrum to specific values, for example 1.0
ppm and 6.0 ppm: sp=1.0p wp=5.0p. For baseline correction
type dc, or bc to reference a peak, for example,
set acetone to 2.05 ppm (in organic
solvent) or 2.225 ppm (in D2O) click as near as possible
to the center of the acetone peak, nl rl(2.225p).
For peak-picking
dpf displays the peak frequencies
th allows
threshold level adjustment with the left mouse button.
9. Data are not normally saved but it possible to save fid in a data
directory. Click Data, then type svf and file name.
10. Plot Data -- Click Main
Menu/ Display/Plot/Plot/Scale/other options/Page
Page is the command that sends the information to the plotter.
Every option selected before page will be included in the plot.
11. Put Lock Standard Back
Take out the sample and insert the lock sample back
into the magnet.
Reset lock for the lock standard.
If somebody is waiting, insert his/or
her sample instead of the lock sample. Make sure the sample is
spinning. In return, the person will log the user off.
12.
Log off the Computer
First close down VNMR and then exit the software. Reverse the
logon procedure.
Tuning -- NOTE: Ask for help of someone who has done
it before! Never tune the probe unless without training.
1. Insert the sample.
2. Setup 1D proton parameters (make sure tn=H1,
dn=C13 and correct solvent). Type su. Now machine
knows CHAN1 for H1 and CHAN2 for C13.
3. Disconnect the cable from the 1H preamplifier and connect
it to the PROBE J5321 port on the TUNE INTERFACE panel.
4. Press the CHAN + button so that the readout is channel 1
for proton. Make sure the red indicator light is not flashing
before proceeding . If it is flashing, check the connections
just made with the two cables.
5. Tune the probe using the appropriate stick (note that the red stick always for proton). The
upper, smoother part is the tune and the lower, rougher part is
the match. As the probe gets closer to being tuned, the
number on the INTERFACE display will decrease.
6. Press the ATTEN buttons until the readout is 8. Continue
tuning the probe until the displayed number is less than 10.
7. If the numbers do not change easily (or no response) when
turning the rod, there must be something wrong. Find someone
who can help. Do not force the tuning rod or turn it too far.
8. Disconnect the tuning function by pressing the CHAN buttons
until the readout is 0 (otherwise acquisition will not be possible).
Reconnect the rf cables to their original positions. The red
indicator light should turn off.
Repeat the process for tuning C13, by choosing CHAN 2, and
C13 cable connections.