 |
Detailed Procedure for Use of
the Spec 20 Spectrometer
|
|
A. Introduction
Spectrophotometry is a technique used to measure how much radiant
energy a substance absorbs at varying wavelengths of light. Pigments
such as chlorophyll and other colored materials (i. e.,
dyes) absorb energy in the visible light range (380-760 nm).
The Spec 20 can not produce data for substances which
absorb at shorter wavelengths (i.e., ultraviolet = UV) or at longer
wavelengths (infrared = IR) The instrument is not capable of a
wider absorption range. The other spectrometers and the Spectronic 21 are used to make measurements in the UV
or IR regions.
By measuring the Absorption Spectrum of a substance, i.e. all the wavelengths at which it absorbs, it is possible to identify
it or at least place it in a particular class of compounds. The
wavelength at which peak absorption occurs, the absorption maximum,
is very useful when trying to identify an unknown. By creating
and measuring a series of standards, it is possible to quantify
the amount or concentration of a substance in a sample. For example,
in their pure form, the nucleic acids can be quantified by absorption
measurements in the UV range.
Wavelength is selected by adjusting a prism within the instrument
such that only a narrow range of light wavelengths are directed
through the sample.
B. Theory: Transmittance and Absorbance
Transmittance (T) is defined as the fraction of incident light
which is transmitted, i. e., passes through, a sample.
Thus, T = I/Io, where Io equals the intensity of light which strikes
the sample and I is the intensity of light after passing through
the sample. Transmittance is usually expressed as a percentage:
%T = (I/Io) x 100
Absorbance (A), or optical density, is a logarithmic function
of T and is expressed as:
A = log10 (1/T) = log10 (Io/I)
Note that absorbance has no units.
So, for example, at 100% transmittance, A = log 1.0 = 0. At
50% transmittance, A = log (1/0.5) = 0.30. The Spec 20 has two
scales, one calibrated from 0 to 100% Transmittance and the other
as Absorbance, ranging from infinity to 0. Note that the highest
calibrated unit of absorbance is 2.0. Spectral data are usually
plotted as absorbance (Y-axis) vs. wavelength or concentration
(X-axis).
C. BLANKS
In order to measure the absorbance of a particular substance
in a reaction mixture, it is necessary to first "zero out"
the spectrophotometer such that only the absorbance of the substance
of interest is measured. This is done with a blank - a cuvette
which contains all the carrier solvents EXCEPT the substance of
interest. A separate BLANK is needed for every unique reaction
mixture.
Formulation of the BLANK depends on the particular experiment
and may involve use of dyes or other colorimetric indicators.
D. How to Operate
the Spec 20:
Zero the Spec
20 to infinite absorbance and a BLANK
1. Turn on the Spec 20 and allow it to warm
up for 5 - 10 minutes (left front knob). Set wavelength using
the dial on top of the Spec 20.
Note: the red light indicates the instrument turned on. The light is also red even if the source light is burned out. Check under the instrument to see if light is visible near the source replacment door. Report any lamp failure to the instructor.

2. Prepare a BLANK cuvette by adding all solvents
EXCEPT the substance to be measured. (Refer to the specific experimental
protocol).
A BLANK is
used to calibrate the Spec 20 so that any absorbance attributable
to the solvent and/or glass cuvette can be compensated. By zeroing
the Spec 20 to the blank, an instrumental subtraction will measure
only the absorbance due to the substance in question.
3. With no sample tube in the holder, adjust
the meter needle to read infinite absorbance (i. e. 0%
transmittance) using the left front knob (i.e. power switch).
4. Using a Kimwipe, wipe off/polish the outside
of the BLANK cuvette to remove greasy finger smudges etc. (Wear
gloves).
Using a wax pencil or Sharpie, make a small
vertical mark at the top of each cuvette for alignment in the
sample holder.

5. Raise the sample cover door and insert the
cuvette such that the line on the cuvette lines up with the line
on the sample holder, close the lid.
NOTE: the plastic lid hinge is the most fragile part of the instrument.

6. Using the right front knob, adjust the meter
needle to read 0.00 absorbance (i. e. 100 % transmittance).
This step is called setting the "full scale".
The spectrophotometer is
now calibrated to this BLANK. If the experiment involves multiple
reaction tube formulations, each one will need its own blank and
the Spec 20 must be re-zeroed for each.
Measuring Absorbance
or Transmittance on the Spec 20
7. Remove BLANK and insert cuvette containing
the sample. Close lid.
8. Read the absorbance (lower scale) OR transmittance
(upper scale) as appropriate for the sample.

9. Repeat for subsequent samples which use
the same BLANK. (SEE NOTE BELOW)
NOTE: When taking several measurements at the same wavelength
over a short time period, no need to re-blank for each. Over longer
times, however, the unit may drift and recalibration to the BLANK
will be necessary. If, however, the wavelength is changed, then
re-zero the instrument. If readings are taken over an extended
period or sharing the instrument, re-zero for each measurement.
E. Beer-Lambert
Law
The Beer-Lambert law describes an important relationship that
exists between absorbance (A) and two sample parameters - solute
concentration (c) and length of the light path (l). Simply put,
the law states that absorbance, A, is directly proportional to
c and l, and is represented by the following equation:
A = εcl
where, ε is a constant and A is the absorbance coefficient.
In biological research, concentration (c) is usually expressed
on a mass/volume basis, e.g., ug/ml, length of light path (l)
in centimeters (usually l=1 cm), and ε, the absorbance coefficient,
is also expressed on the basis of weight. For practical purposes,
the light path is the interior diameter of the cuvette and is
the same for all samples. Therefore, a plot of absorbance vs.
concentration yields a straight line with slope ε. Such
a curve using known concentrations of a pure substance is called
a standard curve. A standard curve is then useful for determining
the concentration of the same substance in solutions of unknown
concentration. By algebraic rearrangement of the above equation
to,
c = A / εl
Concentration can be determined from absorbance alone. Absorption
coefficients for biological molecules can be determined experimentally
or can be found in the literature.