 |
Perkin-Elmer Spectrum One FTIR
Spectrometer
|
|
The Perkin-Elmer Spectrum One FTIR Spectrometer
is capable of data collection over a
wavenumber
range of 370-7800 cm-1. It can be configured to run
in single-beam, ratio, or interferogram modes. The best resolution
is 0.5 cm-1. A mid-range Deuterated triglycine sulfate
infrared detector processes signals with a 68340 integrated chip.
The sample compartment is purgeable.
Instrument are interfaced to Perkin Elmer Spectrum
software, which gives the user the ability to extract both qualitative
and quantitative data from the spectrum, and generate custom reports. See below for specific procedures.
FEATURES.
1) Spectrum One systems have:
a) IR Assistant wizard software
b) IR Expert optimizes spectra
c) Multimedia tutorials
2) Spectrum One systems have:
a)Automatic accessory recognition
and optimization
b) Atmospheric correction
c) Absolute Virtual Instrument
(AVI)
d) Look-Ahead - no background
Background -- FTIR spectrometers
record the interaction of IR radiation with a sample, measuring
the frequencies at which the sample absorbs the radiation and
the intensities of the absorption.
Determining these frequencies allows identification
of the sample's chemical make-up, since chemical functional groups
are known to absorb radiation at specific frequencies. The intensity
of the absorption is related to the concentration of the component.
Intensity and frequency of sample absorption are depicted in a
two-dimensional plot called a spectrum. Intensity is generally
reported in terms of percent transmittance, the amount of light
that passes through it.
In the interferometer the light passes through
a beam splitter, which sends the light in two directions at right
angles. One beam goes to a stationary mirror then back to the
beam splitter. The other goes to a moving mirror. The motion of
the mirror makes the total path length variable versus that taken
by the stationary-mirror beam. When the two meet up again at the
beam splitter, they recombine, but the difference in path lengths
creates constructive and destructive interference pattern called
an interferogram.
The recombined beam passes through the sample.
The sample absorbs all the different wavelengths characteristic
of its spectrum, and this subtracts specific wavelengths from
the interferogram. The detector now reports variation in energy
versus time for all wavelengths simultaneously. A laser beam is
superimposed to provide a reference for the instrument operation.
Procedure for Obtaining FTIR Spectra.
1. The instrument and computer are in "always
on" mode--it is not necessary to turn them on. Look at the
spectrometer indicator panel; there will be all green lights.
Make sure the sample compartment is empty. Depress the right switch
on the monitor to turn it on and move the mouse to wake up the
computer.
If the software is not already running, double
click on the Spectrum icon to start the acquisition
program. When prompted, log in with your group name followed by
the password. The instrument is Spectrum One. Do not activate
IR assistant. Press Return or click okay.
2. Enter date, application, user's name,
and/or professors name in the instrument log book and comment on any
problems. If the software is not already
running, double click on the Spectrum Icon to start the acquisition
program. When prompted, log and enter the password. Do not activate
IR assistant. Press Return or click okay.
3. The program will open and check the hardware.
An opening message is titled Accessory Ready Check. It should
say System Ready for Use. Click the Cancel button. If there is
a spectrum already displayed, click on the Delete icon to clear
the window.
4. Obtain a background spectrum for air.
4.1. Make sure the sample area is clean and
empty.
4.2. Choose Scan from the Instrument menu drop down list.
4.3. A window titled Spectrum One Scan and Instrument Setup will open. It has several pages accessed by clicking on the tabs.
Choose the Sample tab and enter the name background for Name. Next click on the Scan tab and enter Background as the Scan type (under Options in the middle of the page.)
All other settings can be left with their default values.
4.4. Click the Apply button, and then the Scan button.
The window will refresh, and soon the background scan appears,
as it is running. A bar in the lower left corner of the screen
shows the progress of the scan. When the scan is complete, a question
appears on whether to overwrite the old background scan. Answer Overwrite.
4.5. Click the Delete icon to clear the spectrum window.
The background scan is not lost!
5. Obtain the sample spectrum.
4.1. Place the sample in the sample beam either
as a KBr pellet or as a thin film on a KBr cell face.
4.2. Choose Scan from the Instrument menu drop down list.
4.3. The Spectrum One Scan and Instrument Setup window will open.
Choose the Sample tab and enter a filename appropriate
for the sample in the Name line. Fill in the description and comments
as required. Then click the Apply button.
4.4 Monitor the progress of the scan as it runs.
5. Analyze and print data.
5.1. Click on the Peaks icon to automatically
label all peaks. Clicking a second time removes the labels.
[If necessary adjust the threshold from Process and then Peak Table from the menu drop down
list and set a new threshold. Close Peak Table, then click on
peaks twice, (once to remove the old labels and again to show
the new) then click on the window enlargement box (next to the
X in the top right corner of the window frame) to enlarge the
image.]
5.2. Zoom a region of the spectrum by dragging
a box around the region of interest and then double clicking inside
the box. The AutoX and AutoY icons will resize the
spectrum to full scale.
5.3. Click on the Text icon to add text to the spectrum.
Select a new font or drag the text to a new position after it
is written.
5.4. Click on the Print icon to print a copy of the spectrum.
5.5. Click the Delete icon to clear the screen for the
next user, or log out if that is appropriate.
6. Clean up
6.1. Clean the sample area.
6.2. Using
a Kimwipe carefully wipe all the sample off the KBr plates. Place
a few drops of chloroform on each of the plates and carefully
wipe that off as well. Repeat the chloroform wash once more. Replace
the salt plates in their tins and place the tins back into a desiccator.
Remove any pellets from the area.
6.2.
Sign out in the instrument log book and record either 'Ok' or
note any problems.