Procedural
Details for ACD NMR Spec Manager
|
|
Spec Manager is for 1-D NMR. It is an easy, almost-intuitive,
and powerful data processing package and spectral data management
system for various types of spectroscopic and chromatographic
data. SpecManager1 by Advanced Chemistry Development (ADC) is in use on one Windows computer in room 130. It allows processing only NMR spectra (NMR module: FID and Fourier Transformed spectra) and drawing chemical molecules (ChemSketch). After a molecule is drawn, the peaks in the spectrum maybe assigned to atoms in the structure or atoms assigned to peaks. SpecManager will create professional quality reports. Standard OLE capabilities allows direct transfer of spectra, tables structures, and graphics into any Windows document. Other modules are available for FTIR , GC/MS, etc.
Processing FID with
SpecManager
The File Open/Import options open spectra or imports FID's. Open
only functions with files already processed by the Spec Manager.
These files must have the .ESP extension. Many different
types of NMR files can be imported but not the Gemini files until
they have been converted into a DOS-type file with a .DAT extension.
Note Clicking on a Operation toolbar produces a new toolbar appears
as a third row with two options, on the left. First a green check
mark to accept
changes or a red X mark to reject
changes before exiting the current mode. Using these options
allows cycling through various options: Phase Correction, Base
Line Correction, Arithmetic, Peak Picking, Integration, Reference,
Annotation, and Multiplet. Each option has several sub-options.
Processing FID Spectra (for a spectra imported into SpecManager).
Save time by complete all processing of the spectrum before spectrum
analysis, i.e. label, integrate, etc.
1. Select the Icon to do a Fourier transform of the FID
current data set or Apodization, an example.
2 The NMR manager has a very handy feature: Interactive Apodization.
This mode allows interactively application of the following options: Fourier Transform and Phase Correction. This feature
permits interactive and quick adjustment of the Window Functions and phase correction parameters for each spectrum.
After processing a file with VHelper then open/import a FID and click Apodization. Entering the Apodization mode for
the first time produces a new toolbar. This command allows interactively:
Fourier Transform and Phase Correction. This feature permits
interactive and quick adjustment of the Window Functions and
phase correction parameters for any spectrum.
3. The Power Spectrum or Magnitude
Spectrum modes disable some of the processing options which produces
beautiful spectra but is not normally used.
4. There
are several ways to correct a spectrum for non-zero phase:
Auto Phasing Simple, Auto Phasing by Baseline Optimization, Auto
Phasing by Symmetrical Lines, and Manual Phasing, which
allows use of a mouse to optimize phasing interactively. This
mode allows adjusting any phase-correction using the mouse. Click
on the Phase Correction toolbar or choose this command. Dragging
with the left mouse button controls the first phase parameter
(Ph0); dragging with the right mouse button changes the second
phase parameter (Ph1) or enter a new value for Ph0 and/or Ph1
by degrees from the degree windows. Manual phase adjustment helps
to fine-tune the phase-correction obtained using the automatic
methods.
5. Zoom
Control Enlarge the spectrum
in the following ways: Using the Horizontal Zoom tool or Rectangular Zoom tool Both tools have the same principle
of operation Click either of these buttons and drag over the spectral
region to be enlarged. The spectrum display area is split into
two areas. The top area displays the entire non- expanded spectrum
with the shaded area denoting the enlarged region; the bottom
area displays the enlarged view. To see details in different sections
of the spectrum drag or resize the shaded area of the spectrum
in the top area. Or enlarge or reduce the spectrum view in vertical
direction only, using the Vertical Zoom commands or Page Up or Page Down keys on the keyboard. The tools differ in
that the rectangular zoom will magnify the y-axis as well as the
x-axis. Select Zoom In or Manual from the View menu or click In
the dialog box that appears enter the coordinates of the area
and click OK. Note: To cancel the Zoom-In mode and to get back
to the full spectrum view use the button on the General toolbar. Right-clicking on the spectrum enables the Horizontal Zoom tool. CTRL + right-click on the spectrum enables the Rectangular
Zoom tool. Click on the spectrum with the Horizontal Zoom
tool active and the spectrum will be enlarged three times around
the clicked point. Hold down the CTRL key when dragging
with either the Horizontal or Rectangular tool active and the
area is zoomed, symmetrically around the starting point.
Analyzing FT Spectra
1. To label peaks in a spectrum select peak picking: auto
peak picking, picking by setting the peak level and manual peak
picking. Information on each selected peak is placed into the
Table of Peaks .This table can also be displayed on screen and
if values are truncated the column width can be increased.
2. To assign peaks first draw the structure and place it on the
NMR spectrum. Then select Analysis and Assignment either by individual
peaks, multiplets, or region. See Additional
Procedures for drawing molecules with ChemSketch.
3. This mode allows peaks to be labeled peaks. There are
three ways of peak picking: auto peak picking, picking by setting
the peak level and manual peak picking. Information on each selected
peak is placed into the Table of Peaks. Default gives two decimal
places for peak position. The number of significant figures can be increased or decreased under Options/ Preferences/Peak/Numeric
Display. Similarly, the number of significant figures displayed
in a table can modified by opening a Table of Peaks or Integrals, then right clicking anywhere inside the table
to bring up the Edit Box. Select Options and then Integer Field
to change the amount of space for digits and the number of digits
displayed.
4. Annotation allow text to be placed by individual or groups
of peaks. Double click on the annotation to edit text or remove
it.
5. Integration includes a manual integration mode and an automatic
integration function. Moreover, the bias (slope or tilt) of the
integral curves can be adjusted. The value of an integral can
be set after selecting the integral then under options select Reference Integral and set the number, e.g.
of protons to 1, 2, 3.5, 5, etc.
6. Select
set reference point (TMS, etc.) for the currently opened spectrum either manually
or automatically.
7. The Multiplet mode allows definition any spectral region as multiplet and then
defining coupling (J values) constants within it, as well
as convert assigned regions to multiplets. First expand the multiplet
region with the rectangular zoom. Set a multiplet by holding the
left mouse button down while dragging the cursor across the set
of peaks; the multiplet area will be marked. To define individual
J values within each multiplet click the first peak with the left
mouse and then click on the second peak with the left mouse button.
Repeat the process. The tentative J values are displayed after each set of peaks have been
defined and can be listed in a table.
8. Measure Peak to Peak Distance Measures the distance
between any two specified peaks. The area may be measured may
be enlarged using either the Horizontal or Rectangular Zoom tools.
To measure the distance: with this tool active point to the starting
point and click, then move the mouse to another peak of the spectrum
and click again.
9. Distance between two any specified points -- The distance
can be enlarged if required with either the Horizontal or Rectangular
Zoom tools. To measure the distance: with this tool active point
to the starting point and click, then move the mouse to another
point of the spectrum and click again. The results of measuring
are displayed on the bottom status bar:
Reporting and Printing
1. Dialog box allows users to specify the settings for spectrum report
page. The dialog box is displayed when selecting either the Copy
to Report Editor or Print command from the File menu.
Many options include: whole and/or expanded spectra, tables of peaks, integrals, reference, annotations,
multiplets, etc.
2. Copy the active spectrum to a new report page in the ChemSketch
Window. Displays the Report Page Setup dialog box allows specific
settings for the report page. Before using this command, ensure
that the NMR spectrum that copy to the ChemSketch Window is represented
correctly in the window. For example, first insert any structure
or Table of Peaks in the Spectrum Window. Once the spectrum is
copied to the ChemSketch Window, changes in the spectral modes
or settings can not be made, although it is possible to change
the style, size, and location of the NMR picture as a graphical
object using the Objects Panel. Spectra data can be copied into
a Report on a single page or on multiple pages. For instance,
the whole spectra might be copied on the first page with expanded
spectra for multiplets on subsequent pages and another page used
exclusively for tables. All spectra and/or just tables may be imported into any Windows program for presentation or publication.
Multi-spectral Operations
1. Arithmetic
Operations -- This mode allows
arithmetic operations (adding and subtracting) with two spectra
at a time. Arithmetic operations can only be applied to spectra
of the same nucleus type and having the same frequency. However,
spectra can have different number of points and width or can be
shifted relative each other along Y-axis (if this is a case apply
the Synchronize option). Before starting subtracting or
adding spectra make sure that two required spectral windows are
selected and it is recommended to switch to the Tile mode.
Click the button on the Operation toolbar the Arithmetic toolbar
appears. Fix Result of Operation To permanently fix the
results of adding or subtracting, for example, for subsequent
arithmetic operations, click this button. Add Spectra Click
this button the spectrum from the selected spectral window will
be added to the one from the active window. For more details click
here. Subtract Spectra Click this button and one of the
two selected spectra will be subtracted from another (active one).
For more details click here. Coefficient field Both the sum and
difference of two spectra are calculated with the use of k factor
(Sum=A+k*B and Diff=A-k*B, where A is an active spectrum). Change
this factor to obtain the desired result.
Step-by-step procedure: first turn on the New Windows and Tile Modes and under Windows Set Scrolling to a number
greater than two. Open the two files make the first spectrum
active and then use the SHIFT key to select the second spectrum (the file tab will turn light blue). Use the Synchronize button to assure both spectra are synchronized. Click on the Arithmetic
button and then either Addition or Subtraction. Select the 'k'
coefficient, type the number and Enter. Use Page Up or Page Down
to zoom the spectrum and see the peaks disappear or appear. If
the results are not satisfactory click the addition/subtraction
button. Or to accept the results select Fix Results and accept the changes
2. Replace Mode If this mode is active (the corresponding
button is depressed on the General toolbar), any spectra loaded
(open, import or copy from DB) will replace the content of the
active spectral window. Alternatively, select either the Add
mode or the New Window mode.
3. Add Mode If this option is active (the corresponding
button is depressed on the General toolbar), the newly loaded
spectra (opened, imported or transferred from the Database Window)
will be added to the same spectral window. This mode is useful
to compare several spectra.
4. New Window Mode If this mode is active all the spectra
loaded (open, import or copy from database) will be placed into
a separate window.
5. Synchronous Mode Sometimes spectra to be compared belong
to different ranges of X. This command allows representation of
spectra in a form convenient for comparing them. Switch to the
Tile mode and select several windows before applying this option.
As this command is selected or by clicking on the corresponding
button, the X-axis of each selected window will be represented
so that they line up. If necessary, the units of measurements
will be recalculated. If any of the Zoom options are applied (View
menu) with this button active, all the selected spectra will be
zoomed simultaneously.
6. Table of Windows Lists the spectral windows that are
currently opened in the Spectrum window. This table contains useful
information on the spectrum from each window. Click on any portion
of the dialog box to get more information: Use this table to switch
among opened windows: double-click on the line containing the
required spectrum and the corresponding spectrum will become active.
Right-clicking within the table opens the context menu with the
commands that will allows management of open files. To edit the
table representation choose the Options command from the context
menu.
Additional
Procedures -- ChemSketch
ChemSketch
is an integrated software package from ACD for drawing chemical
structures, reactions, schematic diagrams, and designing other
chemistry-related reports and presentations. ChemSketch has the
following features: Structure mode for drawing
chemical structures and calculating their properties as
well as the draw mode for text and graphics processing.
Note: switch between Structure and Draw mode using the buttons
in the top left corner of the ChemSketch Window, in the General toolbar or the Space Bar is a toggle between
the modes.
Molecular Structure
1. Create chemical structures using the Structure toolbar,
Atoms toolbar and References toolbar. Instantly see either the
elemental composition or the formula and molecular weight of the
drawn structure or chemical property of the selected structure
on the status bar. After the structure is drawn it can be imported
into the spectrum window.
2. At the lower left hand corner of the ChemSketch window select
either Spectrum or Chemsketch. Select Spectrum after drawing a
structure and it can be imported by selecting Attach Structure to Current Spectrum or the first choice is to
Switch to the Spectrum Window, without the structure.
3. Assignment This mode allows the assignment peaks in
the spectrum to corresponding atoms in the structure and vice
versa. There are three ways of assigning peaks: auto assignment,
assignment by peak and assignment by region. Note that to make
this mode available first insert a structure. Select the Assignment
button on the Operation Toolbar, the Assignment toolbar appears.
It includes the following selection to allows assignment by peak,
region, or multiplet.
By Peak Choose this command or click the corresponding
button on the Assignment toolbar. Place the mouse pointer over
the peak/atom. Click on the peak/atom when it is highlighted.
Note that the red line stretches to the cursor. Drag from the
peak/atom to the corresponding atom/peak and click to assign it.
Notes: If an atom has been assigned to a peak and if the selected
atom has already been assigned to any other peaks, the ''old''
assignment is deleted and the selected atom is assigned to a new
peak. When assigning a peak to an atom and if the selected
atom has already been assigned to any other peaks, the ''old''
assignment is not deleted so it can be assigned to either one
atom to several peaks. View the assigned peaks and corresponding
numbers of atoms in the Table of Assignments by clicking or choosing
the Show Table of Assignments command from the View menu.
ChemSketch and Draw
Several hints to make using the program a bit simpler.
Reseizing
and Moving a Molecule -- After
a molecule is drawn and if it is not centered or the proper size
first click on the molecule to select select it. A selected molecule
will be enclosed in a box with eight small dark squares. Resize
the molecule by dragging the two arrow cursor from any corner
to make it larger or smaller. Dragging from the sides will increase
the width or length of a molecule, but not proportionality.
To move the molecule to another place select the ALT key select
a 4-arrow curser which allows movement of the whole molecule.
There are two modes in ChemSketch. One for drawing a molecular Structure and one to Draw. The later may be useful
to form equations, add text, or diagrams.
For specific applications
procedures see ADC: ACD software for NMR is described: http://www.acdlabs.com/products/spec_lab/exp_spectra/nmr/
1 A free version of software from Acorn is also on the same
computer.