
|
|
|
1. Turn
on the monitor, the Sun computer and related computers system
are not turned off. The computer and laser printer are turned
on from a single switch on the circuit breaker box. The monitor
is turned off when not in use with the on/off switch on the lower
right front side of the monitor.
4. The
operator must wait three (3) minutes before beginning to
open the software for the instrument. Really 3 minutes,
if you do not wait three minutes the system will open in the Demo
(Demonstration) Mode, a message indicating Demo Mode appears in
the lower right hand corner of the window you must begin again.
You can not run the instrument in Demo Mode.Always return the filament
to the standby position when it is not in use,
to conserve its lifetime.
It is easy to visually check the front-top view port for a filament
glow if you are doubt about the filaments condition. The filament
may automatically revert to the standby condition, during an analysis.
Turn the filament on again and analyze the sample again.
| Mass | Species |
| 69 | -CF3 |
| 131 | -CF(CF2)2 |
| 219 | -(CF2)3CF3 |
| 414 | -(CF2)4-N-(CF2)4 |
| 502 | -CF2-N-((CF2)3-CF3)2 |
Select Calibrate
from the Scope window and under Data select unslope
mode. Begin calibration with mass 69. If the mass 69 peak is not
exactly at 69, then double click the center of the mass peak and
select Apply from the Scope Control window to move
the peak to the correct position1.
Repeat the process for each peak. When calibration is finished
turn off the black valve on the instrument, return the CAL gas
to off the CAL gas in the software and return
filament to Standby. The following is an abridged
version of the Tune Parameters Window.
| File | Status | Off |
|
Standby | |
| Filament, eV |
|
||||
| Gas Control | Off | Cal | GasA | etc. | |
| Source Temp | 0 to 300° C | ||||
|
|
|
||||
The tune status window
show the temperatures of the Source, Expansion Region, Nebulizer
Control, and Nebulizer Tip. The status must not indicate Open
as that means a cable is not connected or the thermocouple is
broken.
11. Open Method from the IonStation
Toolbox. Select File, Load Method, e.g. Select keb.mth
and Load. The following selections are shown
and can be modified.
| Parameter | Units |
| Start Mass | 50 amu |
| End Mass | 400 amu |
| Scan Rate | 400 amu/sec |
| End Scan | 60 Scans |
| End Time | 1 min |
| Algorithm Choice | High Resolution |
| Peak Threshold | 50 adc |
| Peak Width | 16 dacs |
| Scan Delay | 8 msecs |
| Comments | Any ID you need |
The window will display
the TIC (Total Ion Chromatography). Select Load Method
(.mth) and under Prop (method) select Sample and
Apply.. Complete step #12 before beginning to Acquire
data.
12. Before connecting the Waters high-pressure pump to the mass spectrometer:
A. Turn on both speedy2
valves for the two roughing pumps
B. Remove the quick lock from the inlet port.
C. Start the pump with the nebulizer3
outside of the vacuum chamber to inspect the flow of methanol
at 0.5 mL per minute. The on/off switch is at the base of the
pump. The only other control that may need adjustment is the flow
rate, i.e. increase the flow to 1.5 mL per minute. Observe the
flow and if it is satisfactory return it to 0.5 ml per minute.
Inspect the spray to be certain it is continuous before turning
on the helium. The helium pressure is set at 100 psig.
D.
Increase the nebulizer tip temperature to 90° C, while
the nebulizer is outside of the instrument.
E. Remove the vacuum flange and attach the nebulizer to the vacuum
system. Be careful not to bend or break the capillary tubing at
the end of the nebulizer.
F. Leave both speedy valves to the two-roughing pumps open
while the nebulizer is attached to the instrument, as well as
anytime the vacuum system is open. These valves can be closed
only when the vacuum chamber is completely closed.
13. Use a 25-microliter syringe to inject a methanol solution
containing about 10 mg/mL of the pure compound. Make about 4 injections
of about 5 microliters each.
Turn on the
filament, after the samples have been prepared, the injection
syringe has been properly rinsed and is filled with the sample
under the Acquire Window select Start. The data
file will append information to a new data file, e.g. if the file
is named GAscan1. The new name will GAscan1_01.
Click on Stop after the sample has been injected a few
times. The next sample injected will be called GAscan1_02.
Look at the data during and after the injections are complete
to insure the sample shows a signal thousands of times above the
background. Change the filament Status to
Standby once the sample have been nebulized.
14. Do not begin to review the mass spectra until all samples
have been analyzed and the nebulizer has been returned to the
Standby position. Click twice on
the Review icon in the IonStation toolbox.
This option allows you to examine collected data. To open a file
under Review select: File, Load Data, and then select
the proper file name e.g. 1unk_04.scan and then Load
the file Once the file is loaded a window will open the labeled
Chromatogram showing the base line and a peak for each injection.
The peaks must be much higher than the background.
Note: the software
does not allow for additional identification of samples being
analyzed; therefore, it is important to make a hand copy of the
scan number and general name relative to the sample ID. For instance,
1unk-14 must have a handwritten cross reference such as: 14 Jul
-- MW 267, cpd #231a. If files are deleted from a sequence the
higher numbers will be decreased.
15. To view the mass spectra first point the cursor at
the beginning of the area you want to select and depress the left
mouse button and drag the cursor to the end of the area
you wish to obtain mass spectra and release the mouse button.
Normally an area under a maximum is selected. Select a couple
of background points with the SHIFT key held down, as you
point to valleys on either side of the selected mass spectra area.
Select the Mass Spectra button to view the spectra. It
will be labeled with the given name plus the selected scan time
and the points used to subtract background.
In order to view the lines of the spectra select the area of the
mass spectra and under the Chromatography window select Graph
and Quantitation. The data can not be printed from quick
print but from the Review Chromatography print menu after
selection of Item and Quantitation List.
16. The software also allows identification of compounds under
Quick Search. First select a portion of the scan under Review:
Chromatography and obtain a mass spectra for that area. Then highlight
an area of the spectrum that best describes the compound and with
the right mouse button select Quick Search from the drop
down menu.
The search program will find ten most likely known compounds that
may be the unknown compound. Each compound is ranked against a
known relative to the known spectra and then to the difference
between the known and unknown. In addition to the three displayed
spectra the IUPAC compound name, its molecular weight, CAS number,
EPA accession number and normally a structural formula is shown.
17. Print under Review Chromatography has four selections
under Item. Normally the selection is left at Mass Spectra
Window. Other choices are: Chromatographic Window,
Search Window, or Quantitation List.
18.
Shut Down Procedure (Check
to make certain the filament is Standby or Off)
A. Reduce the nebulizer control temperature to about room temperature
25° C.
B. Lower the Expansion Region temperature to about room temperature.
C. Disconnect the solvent interface between the instrument and
the methanol nebulizer.
D. Check the nebulizer tip temperature and turn off solvent flow
through the capillary and the helium.
E. Turn off the pump.
F. Place the o-ring seal and stainless steel flange cover over
the open nebulization port.
G. Manually, turn the speedy valves off for the roughing pumps.
H. Close all windows and exit the software.
I.
Turn the instrument off at the main switch and turn off the monitor.
19. Routine Maintenance
The ion volume needs to be cleaned on a regular basis. When
the injection quality degrades cleaning is required.