The following are general or suggested setting. Every procedure may have completely different settings. The Cary Tab is first.Cary Tab (Select the wavelength and absorbance range)
X Mode Start 800 nm Stop 200 nm (or other desired range for the spectrum)
Y Mode Absorbance Y Minimum -0.02 Y Maximum 1.00 (consider other values depending on sample) Cycle Mode off (e. g. unchecked) Options Tab (Selects the single, dual, scan controls, and overlay options) Beam Mode Dual Beam Scan Controls Simple Speed Medium Display Options Overlay Data Baseline tab Correction Baseline Correction Accessory tabs All accessories off Reports tab Name and Experiment Type Comment Add comments which will appear on the printed spectrum and report. Options Use Data Form on Graph 100% Paper Height Parameters on Peaks All Peaks Peak Labels X & Y Labels Autoconvert none Click on the Peak Information button and make these settings: X & Y Labels on Y Decimals 2 Peak Type Peaks Peak Threshold 0.01 Click OK to return to the Reports tab. Autostore Tab Storage off