 |
Atomic Absorption Flame and
Precautions
|
|
Air-acetylene is the preferred
flame for the determination of approximately
35 elements by atomic absorption. The temperature of the air-acetylene
flame is about 2300 °C. An air-acetylene flame can be used
with all Perkin-Elmer burner heads. The step-by-step procedure is given for the operation of the acetylene/air flame.
For the air-acetylene
flame, the acetylene flow is about 4 liters/minute, or 8.5 cubic
feet/hour, using a heat combustion value of 1450 BTU per cubic
foot. The heat given off would be approximately 12,300 BTU per
hour. Usually, ordinary welding grade acetylene is adequate. However,
welding grade acetylene is supplied dissolved in acetone. As tank
pressure falls, the concentration of acetone entering the flame
increases. This increase may produce erratic results when determining
elements that require a fuel-rich flame or elements with
analytical wavelengths in the lower UV range. Also, acetone passing
into the gas control box may damage valves and tubing. For these
reasons, acetylene tanks should be replaced when the tank pressure
falls below 520 kPa (75 psig). Acetylene tanks should be stored
and operated in a vertical position to prevent liquid acetone
from reaching the cylinder valve. If the welding grade acetylene
is not adequate for atomic absorption, atomic absorption grade
acetylene may be available.
The nitrous oxide-acetylene flame has a maximum
temperature of about 2900 °C and is used for the determination
of elements which form refractory oxides. It is also used to overcome
chemical interferences that may be present in flames of lower
temperature. For the nitrous oxide-acetylene flame, the acetylene
flow is about 14 liters per minute or 30 cubic feet per hour.
Using a heat of combustion value of 1,450 BTU per cubic foot,
the heat given off would be approximately 43,000 BTU per hour.
Light emission from the nitrous oxide-acetylene flame is very
strong at certain wavelengths. This may cause fluctuations in
the analytical results for determinations performed at these wavelengths,
particularly if the lamp emission for the element of interest
is weak. Only the nitrous oxide burner head can be used with the
nitrous oxide-acetylene flame. Ignition procedures and operating
conditions for the nitrous oxide-acetylene flame are given in
the appropriate instrument manual.
Gas Controls
Initial operating conditions, including pressure
and flow settings for fuel and oxidant, are listed in the appropriate
instrument manual. For flame ignition and adjustment procedures,
consult the instrument operating manual. Adjustment of the burner
head relative to the light path of the
instrument is necessary to obtain maximum sensitivity.
The burner height may be adjusted by first
lowering the burner (with the flame off) until the burner head
is well below the light beam and then zeroing the readout. Slowly
raise the burner until the head just intersects the beam (indicated
by a positive reading on the readout). The burner should then
be slowly lowered until the readout returns to zero (indicating
the burner is just below the light beam). With elements that require
a fuel-rich flame, the burner height may have to be optimized
by aspirating a standard solution with the flame on and slowly
lowering the burner until a maximum signal is obtained (recheck
zero after lowering the burner and readjust if necessary). Lateral
and rotational adjustment of the burner must be made with the
flame ignited and while aspirating a standard solution of the
element of interest. The burner is adjusted in both cases to obtain
maximum absorption.
For many elements, the fuel/oxidant ratio must
be adjusted for maximum sensitivity. Aspirate a standard solution
and adjust the Fuel, checking zero after each flow change and
readjusting as necessary. Set up the instrument for the determination
of copper, magnesium, or other elements which have an absorbing
wavelength above 250 nm and whose sensitivity is not dependent
on the fuel/oxidant ratio. Do not select an element requiring
a nitrous oxide-acetylene flame. Aspirate a standard solution
of the element being determined and turn the knurled end cap of
the nebulizer counterclockwise until air bubbles into the solution.
If the flame becomes leaner during this operation, slight adjustment
of the fuel flow may be necessary. While observing the absorbance
reading, slowly turn the knurled end cap of the nebulizer clockwise.
The absorbance reading may go through several peaks as the end
cap is turned. Return the end cap to the peak which gives maximum
absorbance (normally the first peak). After adjustment turn, the
locking ring counterclockwise until it rests against the nebulizer
end cap to prevent accidental misadjustment. Nebulizers do not
normally require readjustment unless a different solvent is used.
Many times it is desirable or necessary to
use solvents other than water.
In selecting an organic solvent, the solubility of the sample,
the miscibility of the solvent with water (for extractions), and
the burning characteristics of the solvent must be considered.
Operation of the Air/Acetylene Flame
Caution
Dangerous
Explosive Mixtures may result if conditions are not correct. Never
use oxygen in the burner.
1. After the spectrophotometer
has been turned on. Verify the flow spoiler or impact bead is
installed in the burner chamber. The suitable burner head is installed,
and that the drain is properly connected. Be certain the burner
head safety interlock pin is correctly installed.
Caution
The escape of fuel gas or fuel gas
mixtures from the drain tube, burner system, the gas control,
or the gas connections may produce an explosive hazard. Never
store the waste vessel in a cabinet or confined space. The operator
must be able to observe the level of liquid at all times.
2. Made certain the
FUEL shut off toggle valve in in the closed position, e.g. parallel to the panel.
3. Set the acetylene cylinder
regulator output pressure at 12 psig.
Caution: Never set the
acetylene cylinder outlet pressure above 15 psig. Changes the
acetylene tank if the tank pressure falls below 75 psig. If the
tank is not changed when as the amount of acetylene is exhausted
then acetone will be brought from the tank into the burner control
and cause damage.
4.
Set the air supply output pressure to between 60 and 100 psig.
5. For a single-slot, 10-cm
burner head turn the oxidant selector valve to AIR. Set the auxiliary
Oxidant Flow valve for a flow meter reading of 50.
6. Adjust the FUEL Flow valve
to bring the flat ball to about 20 on the flow meter.
7. Ignite the flame by pressing
the IGNITE button and holding it until the flame is lit.
8. After ignition, minor adjustments
in the flame conditions may be necessary. Allow several minutes
warmup of the burner system. If the flame is uneven, shut it off
and clean the burner head slot. Check the burner alignment.
9. A fuel-lean (oxidizing)
air-acetylene mixtures gives a blue, hot flame. A fuel-rich (reducing)
air/acetylene mixtures gives a yellow, cool flame. Adjust the
flame to recommended conditions.
AAS Periodic Table
