AAnalysis 800 Atomic Absorption Spectrometer
 

These are Specific Procedures for the Graphite Furnace and Flame operations.

GENERAL INFORMATION: The AAnalyst 800 atomic absorption (AA) spectrometer features a combined flame/graphite furnace. It provides automatic exchange of the atomizers (controlled via computer), whereas previous instruments' had to be changed by hand. This improves reproducibility, saves time, and is easier to use. Solid-state detector technology and simultaneous background correction have quieted background noise and boosted detection sensitivity anywhere from 5 to 20%. Specific protocols are described: General User Guide, Graphite Furance, and the Flame Burner. Users must be familar with the procedures for care and cleaning of the instrument..

Model Spectra AA 800

FEATURES
Real Time Double Beam Spectrometer with Wavelength Range 190 - 870 nm.
Diffraction grating used is dual blazed with ruling density of 1800 lines /mm, dual blazed and RLD 1.6 nm/mm.
Lamps position Automatic Turret with computer controlled lamp selection. Built-in power supply for 4 EDL's.   

 Built-in Atomizer for Both Flame and Graphite Furnace (Stabilized Temperature Platform Furnace conditions).
The Detector is Solid-State Detector.
Total Gas Flow Control System.
 All safety interlocks built-in.
Computer controlled motorized burner adjustments.
 Software control by AA Winlab 32 for the Spectrophotometer and all major accessories including Graphite Furnace, Burner Heads, FIAS, Autosampler etc.
TotalFlow™ gas controls for flame AA and a Transversely Heated Graphite Furnace (THGA) with        longitudinal Zeeman-effect background corrector

Picture 800 AASHigh-efficiency optical system for lowest detection limits At the heart of these advanced AAnalyst systems is an exceptionally efficient optical system featuring a unique solid-state detector with a photoactive surface optimized to provide the highest quantum efficiency in the UV region. Combine our state-of-the-art detector with proven monochromator efficiency and spectrometer throughput, and even elements such as As and Ba can be measured with outstanding signal-to-noise ratios.

Longitudinal Zeeman-effect
background correction provides lowest detection limits With longitudinal Zeeman-effect background correction, the amount of light throughput is doubled by eliminating the need for a polarizer in the optical system. All other commercial Zeeman designs incorporate inefficient polarizers that reduce light throughput and diminish performan800 Optical Pathce. With this unique design, the AAnalyst systems provide the lowest detection levels available. To further improve detection levels and accuracy, the AAnalyst systems also include optimized sampling frequency and interpolated background correction.

THGATHGA furnace provides uniform temperature distribution -- The patented THGA tube used in the AAnalyst systems provides a uniform temperature distribution along its entire length. This eliminates cooler temperatures at the tube ends and removes most interferences. With the THGA tube design, accuracy and sample throughput are improved by reducing the need for the time-consuming standard additions technique.

Consistent temperature control enhances performance -- In conventional furnace systems, the heating rate during atomization depends on the input-line voltage. As voltage varies from day to day, season to season or among laboratory locations, so does the heating rate. The AAnalyst high-performance systems use enhanced power control circuitry to maintain a uniform heating rate, so no matter where a system is located, you can be sure that it provides outstanding, consistent performance.Menu WinLab32

WinLab32 software
Designed with extensive input from laboratory managers and AA users around the world, WinLab32 software provides all the tools and features needed to start running samples quickly and meet the requirements of today’s laboratory. Easy to learn and easy to use The extensive Wizard features of WinLab32 make complex tasks easy with step-by-step instructions and Tool Tips, provides additional information about screen text and entry fields. Status panels display the status of each instrument component for easy monitoring. The Analysis List combines standard, sample and method information into one list, showing the exact order in which the analysis will be run. This list also displays the analysis status at all times and can be printed as a summary at the end of the run. Improved productivity WinLab32 software improves laboratory productivity
by reducing the time required for method development, sample analysis and report generation. Furnace method development is completely automated, helping to optimize the pyrolysis and atomization temperatures as well as sample and modifier volumes. You can create methods, review or reprocess data offline, even add samples anytime during an analysis, without
interrupting the active analysis. Recall Calibration eliminates the need for initial calibration, while Edit Calibration gives you complete control over the quality of your calibration curve before you proceed with QC and sample analysis.


This procedure must be modified for the 800

An example of step by step AAS instructions for an Iron Method illustrating typical operation of the Perkin Elmer 5100 aa:

1.WinLab32 Icon Turn on AAS, computer, and printer.
2. Double click on the Instrument Icon from Windows initial screen.
3. The WinLab32 for AA window will open - do not proceed until the 5100 icon underneath the AAS icon is highlighted.
4. While you're waiting for the icon to light up, open your acetylene tank and air line. Acetylene pressure should be at 14 psi (this is marked on the gauge in black) and air pressure should be about 45-50 psi (neither air pressure nor acetylene pressure should vary much between sessions, but it's important to always check it, especially after changing tanks). The gauge that reads pressure in the tank is the gauge on the right. The gauge on the left reads how much fuel is in the tank. NOTE: when first opened, the acetylene tank, will usually read above 15 psi (red zone). Don't adjust the pressure right away, because 93% of the time, after about 10 minutes, it will have sunk to 14.1 psi always check the pressu800 First Menure after you light the flame. NOTE: the acetylene tank should be replaced when the fuel level drops below 75 psi (using the scale on the inside of the circle on the left hand gauge). Since contaminants sink to the bottom of the tank, it is not advisable to drain a tank to absolute empty .
5. Under "Windows" at the top of the page, scroll down and open the Element Parameter file. This file is your method file; you can't run an element unless you have defined a method.
6. The Element Parameter file will ask you whether you want a previously defined file, or a new one (so it is possible to use the same method for different sample runs). In the Element Parameter file, you have three pages to fill out - Main, Calib, and Options.
7. Main
8. - specify the element (standard wavelength and slit width will automatically appear), select an appropriate sampling time (default is 5 sec), make sure the flame sensor is on, and confirm that the fuel is acetylene and the oxidant is air. "Time average" and "AA" remain as is. Calib
9. (Calibration) - enter the concentration of your standards and how you want to refer to them (i.e. Sl, or standard 1, or whatever). The most important thing to do is to make sure that if, for example, you have a 10 ppm standard, the standa rd units are set at mg/L. To change standard and sample units you just double click on the line. It is possible to have standard units and sample units different. You decide the number of significant digits (up to the machine's precision level of course) by entering "5.0" as a standard, or "5.00", etc. Last, click on Nonlinear to change the type of standard curve that will be drawn; absorbance vs. concentration for Fe is only linear up to about 10 mg/L. Options
10. - highlight the printing options you prefer and type in any comments about your samples. This page prints out at the beginning of your run automatically. Save your Element Parameter file. The computer will always use the Element Parameter file most recently opened when you begin running samples. Close your newly named file.
11. You should be back at the Lab Benchtop. Now you will open four files that you will need to manually run your samples. Under Windows, scroll down and select "Flame Control", "Manual Control", "Display Calibration", and "D isplay Data".
12. OK, but before you begin running samples, you need to adjust the lamp, and then optimize the burner head position, the fuel, and the nebulizer.
13. NOTE:  For Fe, you must use the longer burner head with the BLACK PIN. When running samples that use acetylene and air, you use the longer burner head with the black pin, but for any elements that require the hotter N20 flame, you use the shorter burner head with the red pin. I know from experience that the AAS will not light if you mix up the burner heads!
14. Adjust the lamp by first opening "Adjust Lamp" under Windows (see, there really is logic to all this; before you know it you won't need this sheet!). If you have created your Element Parameter file (as you should have by now) the AAS has already selected the Fe lamp. Your lamp current should read 25 or 30, and your lamp energy should be in the 60's or 70's [in the past lamp energy has fluctuated, and Perkin Elmer maintains that even at a low lamp energy precise sample measurements are possible, but if lamp energy drops dramatically it is probably a sign something is wrong - call Perkin Elmer!]. Now, open the lid covering the lamps, and maximize the absorbance bar by 1) moving the whole lamp forward and backward and 2) turning the two knobs you will see near the
15. back end of the lamp, which adjust the lamp's position.
16. Optimize the vertical position of the burner head by first lowering the burner head with the large right hand knob next to the nebulizer. Then open "Continuous Graphics" under Windows and Autozero by clicking on "Autozero" under < B>Continuous. Raise the burner head until you get a reading, that is, when your .000 goes up, to .004, then .018, etc. Lower the burner head until you have 0.000 again, then lower it one quarter of a turn more.
17. Now it's time to light the AAS. Don't close your Continuous Graphics window, just click on the Flame Control window. After opening your air and acetylene lines (air comes through a line from a compressor next door, acetylene is in a tank) click on the flame icon to light the flame. By the way, the exhaust system should be turned ON now!!
18. Adjust the fuel and oxidant levels by using the Fuel Up/Down and Oxidant Up/Down keys on the keyboard (towards the right on the top line of F keys). For iron, I have found that a fuel of 2 and an oxidant of 8 - 10 works well. Basically, you wa nt a lean flame (meaning the blue line on the burner head at the bottom of the flame is about 1/8" high). You don't want a rich flame, which is the result of a high fuel level.
19. To adjust the burner head horizontally and diagonally, go back to the Continuous Graphics window. While aspirating a standard (you'll probably want to make more of the standard you choose to aspirate, since a lot of liquid can be used up in thi s step) turn the large shiny left hand knob (horizontal) in both directions until you reach maximum absorbance. Repeat this process with the inside-the-compartment-to-the-right-of-the-burner-head shiny knob (diagonal). Some people like to readjust the ver tical knob at this point, to fine tune it. I don't, but try it yourself and see what you prefer.
20. The nebulizer is also adjusted using the Continuous Graphics window. Loosen the lock on the nebulzier by turning the "locking ring" counterclockwise. This will allow you to adjust the sample flow rate by twisting the "knob ring"; the knob ring being the ring which the nebulizer tube directly connects into (the locking ring is immediately behind the knob ring, touching the knob ring in fact when it is completely tightened). When the nebulizer tube is no longer locked (i.e. the locking ring has b een loosened) you can adjust the rate at which the tube sucks samples by twisting the knob ring. Now, put the nebulizer tube in a standard, and turn the circular knob ring until bubbles come OUT of the tube in your standard; then turn the circular knob ring in the other direction until you reach a maximum absorbance on your Continuous Graphics window.
21. NOTE: it is my experience that the nebulizer sucking rate does not change much at all in between running samples. When you have a good rate, I recommend not messing with it unless you're having problems. Once you have adjusted the nebulizer using absorbance a few times, you will get a fe el as to what is a good rate for the solution to drop in the neck of the volumetric flask. NOTE. Remember to always "relock" the locking ring after making an adjustment!!
22. Now you can close the Continuous Graphics window. Then, click on the Manual Sampling window, where you will want to double click on the line above the "save data on/off" icon, to name the file where your sample data will b e stored. If you do not name a file and thereby cause the "save data on/off" icon to be highlighted, your data will not be saved!!!
23. Now you're ready to begin running your standards/samples. You will always begin a sample run by running a standard curve. Simply put the nebulizer tube in your first standard (your "zero" standard") and scroll down under Calibration to autozero . When the READ icon stops being highlighted, that means the AAS is finished reading that standard, so you will remove the tube, wipe it with a ChemWipe, put it in your second standard, and scroll down to your next standard . Repeat this process until all your standards have been analyzed. NOTE: some people think the neb tube should always be placed in distilled water in between samples, to flush the tube. I have tried it with and without the de-ionized water and found that if you let the nebulizer tub e sit in your standard for one second before its analyzed, then you don't need to do the distilled water wash every time. I tell the AAS to delay analyzing the standard/sample after I hit READ by typing in 1.0 second Delay in the Elem ent Parameter file, Main section. Experiment yourself!
24. You have to use the mouse to analyze standards (and the nebulizer tube must be in the standard before you click on autozero, Sl, S2, etc.) but to analyze samples you can either use the mouse to click on the large READ icon on the Manual Cont rol screen or the keyboard (in the middle, top row). After your first standard curve is run, rinse with de-ionized water and then put the nebulizer tube in your first sample, hit READ (mouse or keyboard), and then keep the tube in the sample un til the READ icon is no longer lit up. Rinse with de-ionized water, then put the tube in your next sample, hit READ, wait, etc. I recommend running a standard curve after every 10 to 15 samples. In between every sample, I always wipe the neb ulizer tube with a ChemWipe.
25. The Display Data window will show you the concentration of your most recent sample/standard. The Manual Control window will keep track of the total number of samples/standards you've run. To view your most recent standard curve, click on the Display Calibration window. To edit a standard curve, click on Edit Calibration.
26. When you've finished running all your samples/standards, you turn off the flame (using keyboard or Flame Control window), close the acetylene/air lines, and bleed the gases remaining in the lines (click on bleed gases in Flame Control window).
27. To save your data file in, for example, ASCII, so that you can transfer the information into a spreadsheet, go to the Data Benchtop. Then scroll down to Reformat Data. A window will open that will allow you to call up your data fi le (double click on the data line), and then choose various options for what you want saved in your ASCII file. The options are too numerous to list here; most are pretty straightforward, but for a detailed listing consult the manual. Once you've made your choices, simply click on Execute Reformat.
28. To view your ASCII file, exit the software. At the C prompt, type x. Scroll up using the arrow keys until you find Data - AAS. Hit return twice and you will see a listing of all the Data files. Then follow the commands at t he bottom of the page: hit V to view the document, C to copy it onto a disk, etc.


AAnalyst 800
Atomic Absorption Spectrometers
Specifications
System Design: The AAnalyst™ 800 is a fully-integrated benchtop design atomic absorption
spectrometers, incorporating all spectrometer, flame and graphite furnace components in a single instrument,
offering fully automated exchange of flame and furnace atomizers at the touch of a button.
Optical System Photometer: Real-time double-beam optical system (single-beam for Zeeman furnace operation with the
AAnalyst 800). Front-surfaced, reflecting optics with protective coating. Optical system sealed within protective
cover.
Monochromator: Littrow design with motorized drive for automatic wavelength selection and peaking.
Wavelength range: 190 - 870 nm. Diffraction grating: 1800 lines/mm blazed at 236 nm and 597 nm. Grating area:
64 x 72 mm. Reciprocal linear dispersion: 1.6 nm/mm (nominal). Focal length: 267 mm. Spectral bandwidths:
0.2, 0.7 and 2.0 nm, dual height; motorized slit drive for automatic slit selection.
Detector: Wide-range segmented solid-state detector, including a built-in low-noise CMOS charge amplifier array.
Automatic Lamp Selection: 8-lamp holder with built-in power supplies for hollow cathode and electrodeless
discharge lamps. Computer-controlled lamp selection and alignment via AA WinLab™ software. Lamp elements
and recommended operating currents are automatically recognized and set when using Perkin-Elmer Lumina™
hollow cathode lamps.
Flame System Gas Controls: Fully computer-controlled with oxidant and fuel monitoring. Keyboard-actuated remote ignition
system with air-acetylene. Acetylene flow is automatically adjusted prior to the oxidant change when switching
to or from nitrous oxide-acetylene operation. TotalFlow™ control of the oxidant and fuel gases for constant
fuel:oxidant ratio.
Safety Functions: Interlocks prevent ignition if the proper burner head, the nebulizer/end cap, or the burner drain
system is not correctly installed; the liquid level in the drain vessel is incorrect; or gas pressures are too low.
Interlocks also will automatically shut down burner gases if a flame is not detected, or if any of the other interlock
functions are activated. Provision is included for safe shutdown from all operating modes in the event of a power
failure.
Flame AssmblyBurner System: Premix burner design that can be moved automatically into the sample compartment via software control and a motorized carriage. Alignment of the flame in the light beam is fully automatic, using a motorized
burner mount for vertical and horizontal burner adjustment and automatic software-controlled self-optimization of the burner position. The burner is equipped with a high-strength inert mixing chamber, angled to ensure proper
drainage. Includes adjustable Universal GemTip™ corrosion-resistant nebulizer and an all-titanium, 10-cm, single-slot burner head for air-acetylene operation.
Background AAnalyst 800: Continuum source double-beam background correction using a high-intensity deuterium arc lamp.
Correction AAnalyst 800: Longitudinal AC Zeeman-effect background correction using a modulate 0.8 Tesla magnetic field oriented longitudinal to the optical path. The magnet is automatically switched on during the atomization step only. Rollover detection is built-in. Also built-in is continuum source double-beam background correction for uses with flame operation.

Furnaces AAnalyst 800: Built-in fully computer-controlled Transversely Heated Graphite Atomizer (THGA™). The graphite tube is transversely heated providing a uniform temperature profile over the entire tube. AAnalyst 800: The graphite furnace can be moved automatically into the sample compartment and positioned via software control and a motorized carriage. An external protective gas stream around the graphite tube prevents the entrance of outside air to maximize tube life. An internal purge gas goes through the graphite tube to remove the volatilized matrix vapors during drying and thermal pretreatment. The two gas streams are computer-controlled independently. Pneumatic opening and closing of the furnace for easy tube change.
Common Furnace Program Flexibility: Analytical programs with up to 12 steps can be set up. Each step can be programmed
Features: with the following parameters:
Temperature: Ambient up to 2600°C (up to 3000 °C with AAnalyst 800) in steps of 10 °C.
Ramp Time: 0 to 99 s in steps of 1 s.
Hold Time: 0 to 99 s in steps of 1 s.
Internal Gas Flow: 0 mL/min (gas stop), 50 mL/min (mini-flow), 250 mL/min (full flow); can be switched over
to another type of gas (Alternate Gas).
Furnace Opening and Closing: Pneumatically-operated by software command.

Required Inert Gas: Argon. inlet pressure 300 kPa (3 bar) minimum. Maximum gas consumption is 700 mL/min
with the AAnalyst 800, 1220 mL/min with the AAnalyst 800.
Water Coolant: A circulatory cooling system is included with the AAnalyst 800.
When operating the AAnalyst 800 without the circulatory cooling system, cooling water meeting the following
specifications should be used: Sediment-free drinking water; 20-40 °C; flow rate not less than 2 L/min; pressure
between 2.5 and 4.5 bar; pH between 6.5 and 7.5; hardness not greater than 14°dH or 100 ppm.
Furnace Sampler Table: Installed in front of the furnace unit. Removable sample tray with 88 and 146 sampling
Autosampler positions for sample and reference solutions and 1 overflow container for pipet washing. Minimum sample
requirement: ca. 0.1 mL.
Dispensable Volume: Sample and Reagent: 1...99 mL, selectable in increments of 1 mL.
Max. dispensable Vol. 99 mL (sample volume + reagent volume). Flushing volume 1.3 mL, fixed.
Electronics: The autosampler is powered from the spectrometer and is software-controlled.
Data Control Complete PC control using AA WinLab software operating under the Microsoft® Windows 95® operating
System environment. Provides complete control of the instrument and its major accessories plus data handling and storage.
Data Handling: Instrument readings linear in absorbance (-0.500 A to +2.000 A), concentration or emission
intensity with continuously variable scale expansion from 0.01 to 100 times. Integration times operator-selectable
from 0.1 to 60 sec. in increments of 0.1 sec. Reading modes include time-averaged integration, non-averaged
integration (peak area), and peak height measurement. Includes built-in statistics. Up to fifteen (15) standards and
a choice of proven calibration equations. Reslope of the analytical curve using a single operator-selected
calibration standard. Built-in IEEE-488 interface for computer connection and use of optional accessories.
Minimum PC Configuration: Intel® Pentium® processor; 90 MHz; 16 MB RAM; 1 serial, 1 parallel
(bidirectional) interfaces; 500 MB hard disk drive; 3.5" disk drive; CD-ROM drive; SVGA graphics board and
compatible monitor; mouse pointer; MS Windows 95. A compatible printer is required for hard copy data display.
Dimensions AAnalyst 800: 110 cm wide x 65 cm high x 70 cm deep (104 cm deep with furnace autosampler).
AAnalyst 800: 187 kg (without controller and cooling system).
Power 230 V (+5%/-10%), 50/60 Hz; 5000 VA (AAnalyst 800).
Requirements Electrical Protection: As defined in EN 61010-1-1993 (IEC 1010-1); insulation class I; insulation category
(overvoltage category) II; pollution degree 2. Technical Certification: Designed and tested to be in compliance with the legal requirements for technical instruments Standards including IEC 348 and VDE 0411 and CSA 22.2 No. 151 and the U.S. Federal Communications Commission standards for radio frequency interference. Also complies with German legal requirements for radio interference suppression (better than grade A/0871). The instrument is developed and produced in compliance with ISO 9001.
The AA WinLab software provides required control parameters for GLP and instrument performance validation.
Safety and EMC standards: EN 61010-1-19993 (IEC 1010-1: 1990 + A1, A2, modified). Electromagnetic
compatibility: EN 50 081-1:92 for emission, EN 50 082-1:92 for immunity.
Environmental Ambient temperature: +15 °C to +35 °C. Relative
Requirements humidity: 20 to 80% non-condensing.
Cooling System Self-priming recirculatory system with fan-assisted heat exchanger (standard with the AAnalyst 800) for constant cooling of the graphite furnace. Water temperature during operation approx. 36 °C; water flow 2.5 L/min.
Power requirements: 230 V (+5%/-10%), 50/60 Hz; approx. 140 VA. Dimensions: 20 cm wide x 375 cm high x 50 cm deep,
18 kg with coolant.